Basic QA Ultrasound Phantom
ATS 535-H

Product Description
The Model 535-H rubber-based, tissue-mimicking (TM) phantom is designed to fulfill the basic testing requirements of a Quality Assurance Program. The phantom is designed with a combination of monofilament line targets and six tissue mimicking cylindrical targets of varying sizes. The monofilament line targets have a diameter of 0.12 mm, to optimize the displayed image at frequencies ranging from 2.25 to 7.5 MHz. Four groups of line targets are provided to evaluate the vertical and horizontal calibration measurements, axial-lateral resolution and the dead zone.
***Lifetime Warranty***
Tests Performed
- Dead Zone or Ring-Down
- Vertical Measurement Calibration
- Horizontal Measurement Calibration
- Sensitivity/Penetration
- Functional Resolution
- Focal Zone
- Axial & Lateral Resolution
- Image Uniformity
|
Uses:
- Quality Assurance
- Basic Clinical Performance
- In-Process Manufacturing
- Teaching
|
Specifications
|
General
|
Overall Dimensions
|
23.4 x 20.5 x 9.5 cm
|
|
Weight
|
3.1 Kg
|
|
Scan Surfaces
|
4
|
|
Scan Surface Dimensions
|
7 x 8 cm
19 x 8 cm
|
|
Tissue Mimicking Material
|
Type
|
Urethane Rubber
|
|
Freezing Point
|
< -40°C
|
|
Melting Point
|
> 100°C
|
|
Attenuation Coefficient
|
0.5 dB/cm/MHz ± 5.0% @3.5 Mhz
|
|
Speed of Sound
|
1450 m/s ±1.0% at 23°
|
|
Line Targets
|
Material
|
Monofilament Nylon
|
|
Diameter
|
0.12mm ± 5%
|
|
Position Tolerance
|
± 0.1mm
|
|
Vertical Group
|
|
|
Number of Targets
|
17
|
|
Interval Spacing
|
1.0 cm
|
|
Depth
|
1.0 - 18.0 cm
|
|
Horizontal Group
|
|
|
Number of Targets
|
10
|
|
Number of Groups
|
2
|
|
Interval Spacing
|
2.0 cm
|
Depth
Scanning Surface #1
Scanning Surface #3
|
5.0 cm
5.0 cm
|
|
Dead Zone Group
|
|
|
Number of Targets
|
9
|
|
Lateral Displacement
|
5.0 mm
|
|
Interval Spacing
|
1.0 mm
|
|
Depth
|
2.0 - 10.0 mm
|
|
Axial-Lateral Resolution Group
|
|
|
Number of Targets
|
6
|
|
Lateral Displacement
|
1.0 mm
|
Interval Spacing
Axial Depths
Lateral Depths
|
5.0, 4.0, 3.0, 2.0, 1.0 mm
4.5 & 16.5 cm
5.5 & 11 cm
|
|
Anechoic Target Structures
|
Type
|
Non-echogenic, cylindrical
|
|
Number of Targets
|
6
|
|
Diameters
|
4, 8, 10 mm
|
Interval Spacing
(center to center)
|
1.0 cm
|
|
Depth
|
3, 5, 12, 15 cm
|
|