Contrast Detail Phantom for Mammography
Model-18-252

Product Description
Introduction
The Contrast Detail Phantom for Mammography is designed to provide a means of quantitatively testing and monitoring the total performance of an entire mammographic imaging chain. Its small size, as well as the number and distribution of holes simulating embedded objects, make this phantom particularly useful in evaluating digital spot mammography systems.
With 49 holes generating subtle contrast variations, the phantom makes it possible to detect small changes in overall system performance. The Contrast Detail Phantom for Mammography contains a 7 x 7 matrix of objects. The diameter of each row of objects decreases from 0.169 to 0.007 inch. In each row, the subject contrast decreases from approximately 6.6% to 0.41% at mammographic energies.
Applications
The Contrast Detail Phantom for Mammography is easy to use…Simply place the phantom on the image receptor surface in the same position as a breast. Position the x-ray tube and compression device as in a craniocaudal examination. When using the phantom on prone-position breast biopsy systems, use the rotating top plate of the phantom and the compression device to secure the phantom against the image receptor. Choose the appropriate kV and mAs factors (26 kV and 60 mAs works well on most systems), or select automatic exposure control.
- tal units
- Comparison of image quality with various screen-film combinations
- Evaluation of conventional, as well as digital and stereotactic modalities
- Determination of the optimum exposure technique, e.g., by variation of tube potential
- Comparison of image quality at various object thicknesses, by variation of the amount of Plexiglas® at a fixed density
* Designed by M.A.O. Thijssen, Ph.D., K.R. Bijkerk, MSc. And J.M. Lindeyer, BSc., Project: Quality Assurance in Mammography (QAMAM), Department of Diagnostic Radiology, University Hospital, St. Radboud, Nijmegen, The Netherlands
Key Features
|
Optimized for digital imaging
Easy-to-use, compact and lightweight
Closely simulates scattering conditions of
the breast
Rotatable support plate accommodates
prone-position x-ray units. The plate can be
returned to a position which does not interfere
with placement of the phantom on flat surfaces
Geometrically-increasing hole depths result
in linearly-increasing x-ray transmission
|
 |
Geometrically-increasing hole diameters
enable quantitative measurement of the contrast
threshold of the mammographic system.
Rotatable support plate accommodates
prone-position x-ray units.
|
Product
Specifications
|
|
Object depth
Column 1: 0.033 in
(0.853 mm)
Column 2: 0.021 in
(0.533 mm)
Column 3: 0.013 in
(0.332 mm)
Column 4: 0.008 in
(0.208 mm)
Column 5: 0.005 in
(0.129 mm)
Column 6: 0.003 in
(0.080 mm)
Column 7: 0.002 in
(0.50 mm)
|
Object diameter
Row 1: 0.169 in (4.292 mm)
Row 2: 0.099 in
(2.524 mm)
Row 3: 0.058 in
(1.485 mm)
Row 4: 0.034 in
(0.873 mm)
Row 5: 0.020 in
(0.513 mm)
Row 6: 0.011 in
(0.302 mm)
Row 7: 0.007 in
(0.177 mm)
|
Typical contrast at
mammographic energies
Column 1: 6.60
Column 2: 4.20
Column 3: 2.60
Column 4: 1.70
Column 5: 1.00
Column 6: 0.65
Column 7: 0.41
|
Phantom material:
Plexiglas
Dimensions
2.47 (w) x 2.47 (d) x 2.47 in (t) (6.27 x 6.27 x
6.27 cm)
Weight 1.2 lb
(0.58kg)
|