Peripheral Vascular Doppler Flow Ultrasound Phantom
ATS 524 & 525

Product Description
The Models 524 and 525 tissue mimicking Doppler flow phantoms contain four flow channels simulating superficial vasculature. The simulated vessels are located 15.0 mm below the scan surface. Built-in scanning wells are provided to permit the use of water or a low viscosity gel as acoustic coupling agents.
Lifetime Warranty
|
|
Tests Performed
- Flow Velocity
- Sensitivity
- Maximum Penetration
- Location of Flow
- Demonstration of the effects
of stenosis (Model 525)
Uses:
- Quality Assurance
- Routine Clinical Performance
- In-Process Manufacturing
- Final Q.A. Inspection
- Research & Development
- Teaching
|
| |
|
The differences between the Models is the type of flow channels.
|
| |
| Model 524 |
|
Model 525 |
 |
The Model 524 contains four flow
channels with diameters 2, 4, 6,
& 8 mm, without stenosis.
|
|
The Model 525 contains four 8.0 mm diameter flow channels, with built in stenosis at 0, 50, 75 and 90% areas of occlusion.
|
Specifications
Model 524
General
Overall Dimensions: 22x14x10cm
Scan Surfaces: 1
Scan Surface Dimensions: 17.5 x 9.8 cm
Maximum Fluid Pressure: 15 psi (1.05 Kg/cm)
Connectors: Stainless steel hose barb
Weight: 6.1 Kg
Tissue Mimicking Material: Urethane rubber
Freezing Point: <-40°C
Melting Point: >100°C
Attenuation Coefficient: 0.5 dB/cm/MHz ± 5.0%
Speed of Sound: 1450 m/s ±1.0% at 23°
Flow Channels
Type: Circular
Number of Channels: 4
Diameters: 2, 4, 6, 8 mm
|
Model 525
General
Overall Dimensions: 22x14x10cm
Scan Surfaces: 1
Scan Surface Dimensions: 17.5 x 9.8 cm
Maximum Fluid Pressure: 15 psi (1.05 Kg/cm)
Connectors: Stainless steel hose barb
Weight: 6.1 Kg
Tissue Mimicking Material: Urethane rubber
Freezing Point: <-40°C
Melting Point: >100°C
Attenuation Coefficient: 0.5 dB/cm/MHz ± 5.0%
Speed of Sound: 1450 m/s ±1.0% at 23°
Flow Channels
Type: Circular
Number of Channels: 4
Diameters: 8 mm
Percent of Area Occulsion: 0, 50, 75, & 90%
|
Use in conjunction with the Model 700 Pumping System & Model 707 Doppler Test Fluid.
|
If the user requires depths greater than 15 mm, we recommend the use of our Model 528 scanning wedge. The wedge is constructed of the same tissue mimicking material as the Doppler flow phantoms and provides an additional 50 mm of scanning depth.
To see our full line visit our main website.